ELISA (Enzyme-Linked Immunosorbent Assay) is one of the most commonly used protein quantification techniques in the lab, prized for its high sensitivity and specificity. However, behind its seemingly straightforward protocol lie numerous pitfalls that can lead to experimental failure. Whether it's weak signal, high background, or poor reproducibility, these issues are a major headache.

Today, Dr. Connie will systematically walk you through common problems in ELISA experiments and provide scientific, detailed prevention and solution strategies to help you become an ELISA expert with rock-solid results!

 Poor Standard Curve, Low Fit (R² < 0.99)

The standard curve is the foundation of quantitative analysis. If the standard curve is problematic, the concentration of all samples is unreliable.

Standard points are not linear, the curve is "wiggly," R² value is too low.

1. Improper Standard Reconstitution and Dilution

Cause: Incomplete dissolution of standard powder, or inadequate mixing during serial dilution, leading to inaccurate concentrations.

Prevention: Strictly follow the manufacturer's instructions, using the specified diluent. After reconstitution, briefly centrifuge to ensure all powder is collected at the tube bottom. During serial dilution, always use new pipette tips and ensure thorough mixing at each step by pipetting up and down or vortexing.

2. Imprecise Pipetting Technique

Cause: Using uncalibrated pipettes, or inconsistent pipetting speed/angle, leading to well-to-well volume variations.

Prevention: Calibrate pipettes regularly! Pipette vertically at a steady speed, with the tip touching the liquid surface but not the bottom of the well, avoiding bubble formation.

3. Non-constant Incubation Conditions

Cause: Not using a plate sealer or improper sealing during incubation, leading to edge effects; inconsistent incubation temperature and time.

Prevention: Always seal the plate properly using a high-quality plate sealer. Place the microplate in a thermostatic incubator; avoid "bench-top incubation" on the lab counter to ensure even heating across all wells.

 High Background

High background severely compresses the dynamic range of detection, reduces the signal-to-noise ratio, and makes weak positive samples difficult to interpret.

OD values for the blank or negative control wells are much higher than expected.

1. Insufficient or Overly Vigorous Washing

Cause: Incomplete washing leaves unbound enzyme-conjugated antibodies behind; or excessively harsh washing disrupts the binding of coated antigens and antibodies

Prevention: Effective washing is crucial. Use a multichannel pipette or automated plate washer to ensure each well is completely filled with wash buffer. Always adhere to the recommended soak time (e.g., 30 seconds - 2 minutes) to allow for proper interaction. After soaking, thoroughly remove the buffer by sharply flicking the plate and then firmly tapping it inverted onto a thick stack of absorbent paper to eliminate all residual liquid. Repeat this wash cycle 3-5 times. For manual washing, consistency in filling, soaking, and tapping force across all wells is key to achieving uniform results.

2. Uncomplete Blocking

Cause: Insufficient blocking agent concentration, inadequate blocking time, or inappropriate blocking agent type, leaving non-specific binding sites exposed.

Prevention: Use the recommended blocking agent (e.g., BSA, non-fat dry milk, casein). Ensure blocking proceeds for at least 1-2 hours; if necessary, block overnight at 4°C. After overnight blocking, return the plate to room temperature and wash thoroughly before the next step to remove any potential precipitate from the blocking agent.

3. Excessive Antibody Concentration or Cross-Reactivity

Cause: Suboptimal working concentrations of primary or secondary antibodies leading to non-specific binding.

Prevention: Perform an antibody titration experiment to find the concentration yielding the best signal-to-noise ratio. Verify that the antibodies used do not cross-react with other proteins in the sample.

4. Overdeveloped Substrate or Contaminated Substrate

Cause: MB substrate can slowly turn blue upon exposure to light or heat; or the substrate is contaminated by metal ions or oxidizing agents.

Prevention: Precisely control the substrate development time. Stop the reaction immediately once the highest standard point shows sufficient color development. Store substrate protected from light. Use non-metallic containers for preparation and storage.

  Low Sensitivity, Weak Signal (Weak/No Signal)

The target analyte is present, but no signal is detected – the other extreme.

Positive control shows no signal or very weak signal; standard curve has a low slope.

1. Reagent Degradation or Incorrect Addition/Omission

Cause: Enzyme-conjugated antibody or substrate has lost activity (due to repeated freeze-thaw cycles or expiration); complex protocol leads to skipping a reagent addition step.

Prevention: Bring all reagents to room temperature before starting the assay to avoid condensation dilution. Use a marker to check off completed steps on the plate frame or protocol as you go. Regularly check reagent expiration dates.

2. Sample

Cause: Analyte concentration is outside the detection range (too high or too low); sample contains interfering substances (e.g., hemoglobin from hemolysis, lipids from lipemic samples, proteases).

Prevention: Determine the optimal dilution factor for your sample in a preliminary experiment. For complex samples, perform centrifugation or ultrafiltration to remove interfering substances like cell debris, lipids, or certain proteases, using recommended sample storage buffers. (Note: Centrifugation or sample dilution might affect the concentration or state of the target protein; consider this during experimental design.)

3. Insufficient Incubation Time or Temperature

Cause: Antibody-antigen binding has not reached equilibrium.

Prevention: Ensure each incubation step meets the required time and temperature specified in the instructions. Consider extending incubation times if necessary (requires optimization).

 Poor Reproducibility

High variation between replicate wells, between plates, or between experiments makes results untrustworthy.

Coefficient of Variation (CV) between replicates > 20%.

1. Inconsistent Technique

Cause: Differences in pipetting, washing, and plate tapping force between different operators or even the same operator at different times.

Prevention: Establish a Standard Operating Procedure (SOP) and provide uniform training for all personnel. Ideally, the same person should perform the entire experiment.

2. Improper Mixing of Reagents/Samples

Cause: Concentrated reagents or frozen samples were not mixed thoroughly before use.

Prevention: Gently vortex or invert all liquid reagents and samples to mix before use.

3. Microplate Edge Effects

Cause: Temperature variation between central and edge wells during incubation.

Prevention: Use a thermostatic incubator. During incubation, surround the plate with an empty plate frame to create a more uniform thermal environment. 

4. Incorrect Standard Preparation

Cause: Standard was not equilibrated to room temperature before reconstitution, or the diluent used was not the one specified in the instructions.

Prevention: Strictly follow the manufacturer's instructions for standard preparation. Use only the recommended standard diluent for reconstitution and dilution.

 

The Golden Rules for ELISA Success

Plan Ahead: Read the entire protocol before starting. Prepare all reagents and instruments. Design the plate layout.

Precision is Key: Use calibrated pipettes. Follow techniques meticulously. Keep strict timing.

Equilibrate Thoroughly: Bring ALL reagents and samples to room temperature.

Wash, Wash, Wash: This is the lifeline of a successful ELISA!

Control Development: Monitor the reaction closely. Stop it promptly.

Controls are Non-Negotiable: Blank, negative control, positive control – don't skip any!

Catalog

Product Name

KAK13901

Anti-Semaglutide ELISA Kit

RHK13902

Anti-Semaglutide (GLP-1 analogue) Antibody (SAb2275)

KAF86301

Anti-Benralizumab ELISA Kit

KAB76601

Anti-Teduglutide ELISA Kit

KDV03503

Ormutivimab ELISA Kit

KDG17602

Ligufalimab ELISA Kit

KDJ04006

Ifinatamab ELISA Kit

KDK13902

Tirzepatide (LY3298176) ELISA Kit

KAK13902

Anti-Tirzepatide (LY3298176) Human IgG ELISA Kit

KHC33701

Human sCD14 ELISA Kit

KHC15801

Human IL6 ELISA Kit

Terms of sale Website terms of use Cookie policy Privacy
Copyright © 2025 AntibodySystem SAS. All Rights Reserved.            All Products are for Research Use Only