Catalog No.
KAV00301
Description
PRINCIPLE OF THE ASSAY This assay employs the quantitative indirect enzyme immunoassay technique. Recombinant SARS-CoV-2 RBD (BA.5) has been pre-coated onto a microplate. Standards or samples are pipetted into the wells and any Anti-SARS-CoV-2 RBD (BA.5) Human IgG present is bound by the immobilized protein. After washing away any unbound substances, a biotin-labeled Mouse Anti-Human IgG antibody is added to the wells. After washing away any unbound substances, Streptavidin-HRP is added to the wells. Following a wash to remove any unbound enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of Anti-SARS-CoV-2 RBD (BA.5) Human IgG bound in the initial step. The color development is stopped and the intensity of the color is measured.
Specificity
Anti-SARS-CoV-2 RBD (BA.5) Human IgG Antibody
Applications
Used for the quantitative determination of Anti-SARS-CoV-2 RBD (BA.5) Human IgG concentration in serum and plasma.
Detection method
Colorimetric
Sample type
Plasma, Serum
Assay type
Quantitative
Range
1.72 - 110 ng/mL
Sensitivity
0.38 ng/mL
Precision
Intra-Assay Precision (Precision within an assay): <10%
Three samples of known concentration were tested sixteen times on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays): <15%
Three samples of known concentration were tested in twenty four separate assays to assess inter-assay precision.
|
Intra-Assay Precision |
Inter-Assay Precision |
||||
Sample |
1 |
2 |
3 |
1 |
2 |
3 |
n |
16 |
16 |
16 |
24 |
24 |
24 |
Mean (ng/mL) |
34.9 |
11.1 |
3.9 |
36.6 |
11.3 |
4.0 |
Standard deviation |
2.6 |
0.5 |
0.3 |
2.3 |
0.9 |
0.2 |
CV (%) |
7.4 |
4.5 |
7.7 |
6.2 |
7.7 |
5.5 |
Recovery
80-120%
Shipping
2-8 ℃
Stability and Storage
When the kit was stored at the recommended temperature for 6 months, the signal intensity decreased by less than 20%.